IntelliStart Sample Tune and Develop Method (STDM) fails to find parent ion, even though a strong signal is seen in the Tune page and with manual acquisition from the Tune page. - WKB10751

Article number: 10751

SYMPTOMS

  • IntelliStart STDM fails to find parent ion
  • Signal on Tune page is strong
  • Signal detected on manual acquisition from Tune page is strong
  • Correct molecular weight entered in the STDM setup
  • Correct bottle position selected in STDM setup
  • Sample was purged/primed several times

ENVIRONMENT

  • Xevo TQ-S
  • MassLynx 4.1
  • IntelliStart: Sample Tune and Develop Method (STDM)

CAUSE

Unknown

FIX or WORKAROUND

  1. Close MassLynx and the ACQUITY MS console.
  2. Reboot the MassLynx acquisition PC.
  3. Reboot the EPC.

ADDITIONAL INFORMATION

 

id10751, MLYNX, MLYNXV41, SUPMM, XEVOTQS, XEVOTQSIVD, XEVTQSMIVD