QDa poor signal - WKB107391

Article number: 107391

SYMPTOMS

  • Poor signal on Acquity QDa ,
    • System includes a tee to split flow between QDa and a second detector        
    • Signal good on second detector

ENVIRONMENT

  • ACQUITY QDa Mass Detector

CAUSE

QDa probe partially blocked sending flow to second detector

FIX or WORKAROUND

  1. Replace QDa probe

ADDITIONAL INFORMATION