Vion IMS QTof does not pump down after venting - WKB1055
Article number: 1055
SYMPTOMS
- Turbo speed stays at 0 when pumping down
ENVIRONMENT
- Vion IMS QTof
CAUSE
- Turbo controllers in an error state
FIX or WORKAROUND
- Turn off electronics.
- Unplug the mains cable for 30 seconds.
- Reconnect mains cable.
- Turn on electronics.
- Pump down instrument.
ADDITIONAL INFORMATION
- Turbo controllers are reset when all residual current is drained from the instrument.