How to power cycle a Device+ - WKB324168

Article number: 324168

OBJECTIVE or GOAL

Power cycle a Device+.

ENVIRONMENT

  • Andrew+
  • OneLab
  • Device+

PROCEDURE

  1. Press and hold the Power button for 5 seconds, until the LED indicator begins blinking white.
  2. When the LED indicator begins blinking white, release the Power button. The Device+ automatically shuts down.
  3. Disconnect the power cord from the Device+.
  4. Wait for at least 10 seconds after disconnecting the power cord.
  5. Reconnect the power cord to the Device+. The unit should power-on automatically.
  6. The blinking white LED indicates that the Device+ is initializing.
  7. The Device+ is ready for use when the LED turns green.