Lost connection between Andrew+ and pipettes - WKB284362

Article number: 284362

SYMPTOMS

  • Single-channel pipette does not connect
  • Power-cycling the pipette and robot does not help

ENVIRONMENT

  • Andrew+
  • Pipette+

CAUSE

Unknown

FIX or WORKAROUND

  1. Fix pipette connection problems | Andrew Alliance Help Center

If you’re having trouble connecting or reconnecting your pipette to your Andrew+ robot or Pipette+ stand, try the steps below.

How to reconnect your pipette to the robot/stand?

  1. Power-off the pipette by pressing the power button for 10 seconds.

undefined

NOTE: The 10-second press forces the pipette into a hard restart.

2. Power back on the pipette and wait until the message "Hello, Please Wait..." disappears.

3.Place the pipette back on the tool rack/stand.

4. Try again to execute the same protocol.

If the issue persists:

  • Reboot the pipette as in step 1 and wait 1 minute. Then execute the same protocol again.

  • If the issue is still present, power-cycle Andrew+/Pipette+:

    • Turn it off by pressing the main button.

    • Remove the power cable from Andrew+/Pipette+.

    • Wait 10 seconds.

    • Plug the cable back in and wait until the LED is green.

    • Execute the same protocol again.

If the above actions do not solve the issue, collect the log files of the unit for further analysis.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.