"Sample Fluidics over pressure" error for ACQUITY UPLC Sample Manager - WKB99455
Article number: 99455
SYMPTOMS
- Instrument does not complete initialization
- Console shows "Sample Fluidics over pressure" error
ENVIRONMENT
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC I-Class SM-FL
- ACQUITY UPLC I-Class PLUS SM-FL
CAUSE
Clog in the sample loop
FIX or WORKAROUND
Replace the sample loop.
OR
Flush the sample loop to force any clogs out by connecting it directly to the solvent manager or by sonicating it.