Needle wash solvent consumption is too high with Agilent 1290 Infinity Vialsampler - WKB98715
Article number: 98715
SYMPTOMS
- Needle wash solvent consumption is too high
- Outlet tubing on the peristaltic pump is dropping even if the pump is off
- System had run through about 900 mL of needle wash overnight
- Needle wash solvent consumption zero (the three wheels inside are not drilling)
ENVIRONMENT
- Agilent 1290 Vialsampler G7129B
- G7129A 1260 Infinity II Vialsampler VL
- Agilent G7167B
CAUSE
Faulty peristaltic pump
FIX or WORKAROUND
Replace the peristaltic pump (part 5065-4445).