Pressure fluctuations on ACQUITY H-Class QSM - WKB97153
Article number: 97153
SYMPTOMS
- Pressure fluctuates
- Some channels prime okay but others do not

ENVIRONMENT
- ACQUITY H-Class Quaternary Solvent Manager (QSM)
CAUSE
Clogged GPV filter

FIX or WORKAROUND
Replace GPV filters.
ADDITIONAL INFORMATION
You can also replace the locking rings if necessary: 700005259 - Fitting & Lock Ring, GPV Filter, 4/pk