Pressure fluctuations on ACQUITY H-Class QSM - WKB97153
Article number: 97153
SYMPTOMS
- Pressure fluctuates
- Some channels prime okay but others do not
ENVIRONMENT
- ACQUITY H-Class Quaternary Solvent Manager (QSM)
CAUSE
Clogged GPV filter
FIX or WORKAROUND
Replace GPV filters.
ADDITIONAL INFORMATION
You can also replace the locking rings if necessary: 700005259 - Fitting & Lock Ring, GPV Filter, 4/pk
id97153, A-30SDM, gradient proportioning valve, pressure ripple, UPAQSM, UPPQSM, UPQSM, UPQSM, UPQSM+, UPQSM+, UPQSMBIO, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO