515 HPLC Pump - Fitting broken off in reference valve - WKB96149

Article number: 96149

SYMPTOMS

  • Nut snapped in reference valve.
  • Broken part of nut cannot be removed.

clipboard_eea4992910a1b5af76b415255d0092525.png

ENVIRONMENT

  • 515 HPLC Pump
  • 600 HPLC Pump
  • 1515 Isocratic HPLC Pump
  • 1525 Binary HPLC Pump

CAUSE

  • Fitting over tightened.
  • Fitting damaged.

FIX or WORKAROUND

  1. Replace reference valve (WAT026592).
  2. Verify functionality with ramp test to reference valve outlet.

ADDITIONAL INFORMATION