ACQUITY UPLC Sample/Column Manager - Fitting broken off in valve cartridge - WKB95967

Article number: 95967

SYMPTOMS

  • Screw/fitting snapped in injection valve cartridge.
  • Cannot remove broken part of screw/fitting.

clipboard_e322b555743394c30f71cc3af0f8535d6.png

ENVIRONMENT

  • ACQUITY UPLC Sample Manager
  • ACQUITY UPLC H-Class
  • ACQUITY Arc Sample Manager FTN-R
  • nano ACQUITY
  • Column Managers

CAUSE

  • Fitting over-tightened.
  • Fitting damaged.

FIX or WORKAROUND

  1. Replace injection valve cartridge.
  2. Verify functionality with leak test to column.

ADDITIONAL INFORMATION

PN for H-Class and ARC:700005236

PN for UPLC: 700002765

PN for Nano Acquity TVM: 700004601

PN for column manager : 700003145