Carryover seen on Alliance 2695 - WKB95668
Article number: 95668
SYMPTOMS
- Carryover on Alliance.
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
Bad tubing from seal pack to column.
FIX or WORKAROUND
- Replace tubing.
- 430000922 ASSY, TUBE, PEEK, .062 X .005 X 20 L
ADDITIONAL INFORMATION