Article number: 95650

SYMPTOMS

  • Carryover observed.

ENVIRONMENT

  • ACQUITY UPLC Sample Manager

CAUSE

Wash solvents not washing for long enough in instrument method.

FIX or WORKAROUND

Increase the volume of wash solvent used in the instrument method editor.

ADDITIONAL INFORMATION