Preheater temperature range error (temperature figure) - WKB95640
Article number: 95640
SYMPTOMS
- Preheater temperature range error when running samples with a temperature in brackets.
- Numerous preheaters configured, but no detected errors in the log.
- Active preheater is set to active and produces an error on injection.
- Randomly, an injection increases/decreases from set point and on next injection, produces an error, but temperature returns to set point.
ENVIRONMENT
- ACQUITY UPLC Column Heater
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
CAUSE
Active preheater.
FIX or WORKAROUND
- Replace active preheater and set temperature.
- Run an injection to ensure that the system is working on injection.
ADDITIONAL INFORMATION
Also replace Clip, Retainer (# 700010511) if needed.
If issue persists, active preheater PCB (700005509) is likely the cause.
Active preheater part numbers are 205000756, 205000775, and 205000777 for H-Class Bio systems that are not PLUS systems.
H-Class
- APH, SS, 23.5-in
(205001759) - APH, MP35N, 23.5-in
(205001756) - APH, SS, 12.5-in
(205001774) - APH, MP35N, 12.5-in
(205001754)
I-Class
- ACQ I-Class APH, SS, 12.5 LG, .003 ID (205000894)
- APH, SS, 23.5-in (205001759)