Retention time shift on an ACQUITY H-Class - WKB95183

Article number: 95183

SYMPTOMS

  • Retention time shift to the right
  • Peaks are co-eluting
  • Pressure drops from 12500 psi to 10000 psi
  • Replacing column and guard column does not help

ENVIRONMENT

  • ACQUITY UPLC H-Class

CAUSE

Leaking primary pump head

FIX or WORKAROUND

Rebuild the primary pump head.

ADDITIONAL INFORMATION

For rebuild kit part numbers, see the article What are the part numbers for ACQUITY pumphead rebuild kits? - WKB5426