ACQUITY UPLC H-Class peak tailing and splitting - WKB95105

Article number: 95105

SYMPTOMS

  • Peak tailing and splitting

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC
  • ACQUITY UPLC I-Class

CAUSE

Leak at fitting on tubing between column and detector.

FIX or WORKAROUND

  1. Tighten fitting and/or replace tubing.