"System Idle - Instrument Failure W15xx" error when flow is applied on a 1515 Isocratic HPLC Pump - WKB94894
Article number: 94894
SYMPTOMS
- "System Idle - Instrument Failure W15xx" error when flow is applied.
- System goes quickly to >4000 psi.
ENVIRONMENT
- 1515 Isocratic HPLC Pump
CAUSE
- Overpressure causing the system to shut down.
- Clog in line found from Manual Injector to Column (port 3 on the back of the manual injector).
FIX or WORKAROUND
- Back flush tubing to release the clog.
- Refit tubing.
- Confirm that flow can be applied without overpressure error.
ADDITIONAL INFORMATION
If the clog cannot be flushed out, replace manual injector to column tubing (700001166)
id94894, 1515, 1525, 1525U, 1525UIVD