FTN Sample syringe leak test fails - WKB94706

Article number: 94706

SYMPTOMS

  • FTN Sample syringe leak test fails
  • Replacing syringe does not help

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC I-Class Bio
  • ACQUITY UPLC I-Class Plus
  • ACQUITY Arc
  • ACQUITY Arc Sample Manager FTN-R
  • ACQUITY Arc Sample Manager FTN-R Bio

CAUSE

Syringe metering Valve failure

Syringe metering Drive failure

FIX or WORKAROUND

Replace the Syringe metering valve first prior to ordering critical spare part.

- part number 700011790 SYRINGE VALVE METERING

Replace the Syringe drive assembly

  • part number 700012509 Metering Syringe Drive, High Ratio

  • part number 700012510 Metering Syringe Drive, High Ratio (Bio)

  • part number 700010716 Arc, Syringe Drive 1/4-20

  • part number 700011685 Arc, Syringe Drive 1/4-20 (Bio)

ADDITIONAL INFORMATION

The following part numbers are obsolete

700005201 SYRINGE DRIVE, METERING, HIGH RATIO

700003743 ASSY, SYRINGE VALVE W/COUPLING

Not able to find a solution? Click here to request help.