Broken Flexiglass causes Zp axis home h/w fault - WKB94602
Article number: 94602
SYMPTOMS
- Zp axis home h/w fault errors
ENVIRONMENT
- ACQUITY UPLC Sample Manager
CAUSE
Flexiglass on the top of the XYZ shattered, causing needle to snag and not move freely.
FIX or WORKAROUND
Replace the Flexiglass.
The clear plastic screen is part 405002926 (BRACKET, SAMPLE NEEDLE GUIDE).

ADDITIONAL INFORMATION
The insulation that goes on top of this plastic cover is part 401000404.