Sample syringe H/W lost steps FTN - WKB93912
Article number: 93912
SYMPTOMS
- Error: Sample syringe H/W lost steps
- Error appears at first injection after Purge
ENVIRONMENT
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY APC Sample Manager - FTN
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
- Waste valve is clogged
FIX or WORKAROUND
- Replace Inline Waste Valve (410003180)
ADDITIONAL INFORMATION
Since the Arc does not have a sample pressure transducer there is no error message for high sample pressure. This usually would be the case for an H-Class or I-Class SM-FTN but in rare occasions the error will be for lost steps instead
