Article number: 93912

SYMPTOMS

  • Error: Sample syringe H/W lost steps
  • Error appears at first injection after Purge

ENVIRONMENT

  • ACQUITY Arc Sample Manager FTN-R
  • ACQUITY APC Sample Manager - FTN
  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

  • Waste valve is clogged

FIX or WORKAROUND

  1. Replace Inline Waste Valve (410003180)

ADDITIONAL INFORMATION

Since the Arc does not have a sample pressure transducer there is no error message for high sample pressure. This usually would be the case for an H-Class or I-Class SM-FTN but in rare occasions the error will be for lost steps instead

id93912, UPAQSM, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARHSMFTN, UPARQSM, UPARSMFTN, UPPDAARCB, UPPSMFL