Skip to main content
Waters

How to configure an externally controlled RS232 switching valve with a Column Manager (CM-A or CM-30S) - WKB9338

Article number: 9338

OBJECTIVE or GOAL

Set up an external valve in the console software and in Empower

  • Minimum requirement: Control released with the June 2011 Driver Pack for ACQUITY systems.
    • CM-A supports control for up to three external valves.

ENVIRONMENT

  • H-Class Column Manager - Active UPC2 30cm Single Zone Column Manager Empower Instrument Method Editor.
  • Optional External Valves (RS232-controlled):
    • 417000138 - VALVE, SELECTOR, 2 POSITION, 10 PORT, TI
    • 417000139 - VALVE, SELECTOR, 6 POS., 7 PORT PEEK
    • 417000150 - VALVE, SELECTOR, 10-POS, SS, RS232.

PROCEDURE

ACQUITY Console Setup

  1. Install the external controlled RS232 switching valve with CM-A or CM-30S. Power-on the Column Manager. In the ACQUITY Console, from the system tree, select Column Manager. In the Column Manager information window, click Control > Reset CM (to reset the Column Manager). In the Column Manager information window, click Configure > Operating Mode. Column Manager Select Operating Mode.jpg In the Operating Mode Configuration dialog box, under “Select valve configuration”, select the check box for External 1, External 2, or External 3. Column Manager Column Selection External.jpg
      Note: Select as many valves as are installed. For each external valve, set the appropriate positions for the specific valve type. Click OK, and then reboot the Column Manager when prompted. Column Manager Instrument Configuration Changed.jpg Re-establish communications with the Column Manager and control external valve positions.
    • Illustrated is a Column Manager with two columns, both set at 40 °C, and the internal Column Manager Valve selected for “Column 1” and the external valve selected for “Position 1”. Column Manager Console External Valve Control.jpg
    Click on any of the blue selections to manually make changes.

Empower Instrument Method Editor Setup

Requirement: If the instrument method does not include console-controlled positions or parameters, when you start a method, only the conditions for the Column Manager in that method will be activated.

  1. Example instrument method with external valve controls:
    • Column Manager in “Column Selection" mode Active Preheater "Enabled" Column Selection External Valves (timed events for valve programming not available in column selection mode)
    Instrument Method External Valve Control.jpg Select which position to use with the external valve. Instrument Method External Valve Selection.jpg

ADDITIONAL INFORMATION

The “Advanced” mode of operation enables the “Events” table in the instrument method and allows you to program changes to the valve state during a run.  Positioning of the internal valves changes to get the correct column selected.

id9338, A-30MCM, APH, comms, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP, UPACM30S, UPACMS, UPC2BSM, UPC2CM30S, UPC2CMA, UPC2CNVMGR, UPC2SMFL, UPCMA, UPCMAIVD, UPPCMA, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO

Not able to find a solution? Click here to request help.