Missing peaks and poor chromatography on an I-Class system - WKB92377

Article number: 92377

SYMPTOMS

  • Target peak retention time elutes too late
  • Resolution between peaks is poor and late peaks are missing

ENVIRONMENT

  • ACQUITY UPLC I-Class

CAUSE

The flow cell is leaking inside the IN connection port.

FIX or WORKAROUND

Replace the flow cell: 205015036 - ACQUITY PDA, Flowcell, 10mm 500nL (Standard)

ADDITIONAL INFORMATION

Use the part locator for alternatively sized flow cells.