Missing peaks and poor chromatography on an I-Class system - WKB92377
Article number: 92377
SYMPTOMS
- Target peak retention time elutes too late
- Resolution between peaks is poor and late peaks are missing
ENVIRONMENT
- ACQUITY UPLC I-Class
CAUSE
The flow cell is leaking inside the IN connection port.
FIX or WORKAROUND
Replace the flow cell: 205015036 - ACQUITY PDA, Flowcell, 10mm 500nL (Standard)
ADDITIONAL INFORMATION
Use the part locator for alternatively sized flow cells.
