2414 RI Detector error: Rtd missing or defective out rtd range - WKB9229

Article number: 9229

SYMPTOMS

  • Error: "Rtd missing or defective out rtd range"

ENVIRONMENT

  • 2414 Refractive Index Detector

CAUSE

Oxidized contact in the connector between the 2414 and the oven

FIX or WORKAROUND

Clean the contacts with a contact cleaner.

ADDITIONAL INFORMATION

 