Vials are knocked out of the holder on H-Class FTN - WKB92143

Article number: 92143

SYMPTOMS

  • Vials are knocked out of the holder

ENVIRONMENT

  • ACQUITY H-Class Flow Through Needle

CAUSE

The needle tubing loop is out of position.

FIX or WORKAROUND

  1. Remove the needle.
  2. Reinstall the needle correctly.
  3. Ensure that any extension loop is positioned at the injection valve.

ADDITIONAL INFORMATION

IMG_20200306_073423.jpg

id92143, UPARHSMFTN