Error: "Preheater temperature sensor h/w fault" - WKB92139

Article number: 92139

SYMPTOMS

  • In the ACQUITY Console, the following error appears: "Preheater temperature sensor h/w fault" when trying to set temperature at 50 degrees C.
  • Column active preheater over-temp switch h/w fault
  • Reset does not resolve this.

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY UPLC Column Heater
  • ACQUITY UPLC Column Manager

CAUSE

Unknown.

FIX or WORKAROUND

  1. Reseat the active preheater (APH). 
  2. Power-off SM-FTN and reseat external cable from CH-A to SM-FTN.
  3. Power-on SM-FTN.
  4. Set temperature.

ADDITIONAL INFORMATION