Error: "Preheater temperature sensor h/w fault" - WKB92139
Article number: 92139
SYMPTOMS
- In the ACQUITY Console, the following error appears: "Preheater temperature sensor h/w fault" when trying to set temperature at 50 degrees C.
- Column active preheater over-temp switch h/w fault
- Reset does not resolve this.
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC Column Heater
- ACQUITY UPLC Column Manager
CAUSE
Unknown.
FIX or WORKAROUND
- Reseat the active preheater (APH).
- Power-off SM-FTN and reseat external cable from CH-A to SM-FTN.
- Power-on SM-FTN.
- Set temperature.