No hydrophobic peptides on nanoACQUITY - WKB9023

Article number: 9023

SYMPTOMS

  • No hydrophobic peptides on nanoACQUITY
  • Hydrophilic peptides are showing up
  • Hydrophobic peptides are there if you do a direct injection

ENVIRONMENT

  • nanoACQUITY

CAUSE

Bad trap column is not absorbing the sample.

FIX or WORKAROUND

Replace the trap column with a new one. 

ADDITIONAL INFORMATION