2695 Separations Module: If an error occurs in the compression check and the test is passed when performing again, is it OK to use as usual? - WKB90007

Article number: 90007

ENVIRONMENT

  • 2695 Separations Module
  • Alliance HPLC System

ANSWER

Yes. No problem.
Since the injector purge is performed once before the compression check, air may have been removed at that time.

Not able to find a solution? Click here to request help.