Skip to main content
Waters

Leak from base of Alliance - WKB89999

Article number: 89999

SYMPTOMS

  • Liquid pools below the Alliance

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

The GPV-to-primary tube is leaking at the GPV, and the solvent tray waste line is blocked.

FIX or WORKAROUND

  1. Tighten the fitting on the GPV.
  2. Resolve the waste line blockage.
  3. Replace GPV to primary tubing (WAT270986) if necessary

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.