2998 Photodiode Array Detector: 2998 Photodiode Array Detector spectrum shows a straight line that did not exist before - WKB89776
Article number: 89776
SYMPTOMS
- 2998 Photodiode Array Detector spectrum shows a straight line that did not exist before
ENVIRONMENT
- 2695 Separations Module
- 2998 Photodiode Array Detector
- Empower
CAUSE
Unknown
FIX or WORKAROUND
- Clean the detector
(1) Remove the column, and directly connect the injector outlet of 2695 and 2998 with a union.
(2) Run water at 1 mL/min for about 1 hour.
(3) Run 1:1 isopropanol/acetonitrile for at least 1 hour.
(4) Run water again.
Because the maximum pressure rating of the 2998 detector is 1000 psi, the flow rate in step 2 should be set so that the pressure limit is not exceeded.
After cleaning, run again to check for improvement in the spectrum. - Absorbance may be too low, causing baseline fluctuations.
Extract the chromatogram at the wavelength where the spectrum is linear and compare with when there was no problem. - To determine if this is a instrument problem, remove the column and run water to see if it shows a similar shape.
- Perform 3D Blank Subtraction when processing so that the straight line disappears.
ADDITIONAL INFORMATION
