Peak tailing with ACQUITY UPLC and H-Class - WKB89281

Article number: 89281

SYMPTOMS

  • Peak tailing.

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class
  • ACQUITY inline filter

CAUSE

The ferrule fixing position on the inline filter is shallow, and the dead volume occurs at the column inlet.

FIX or WORKAROUND

Attach a new in-line filter to the column.

ADDITIONAL INFORMATION

In-line filter part number: WKB2796

Installing the in-line filter: WKB72237

Not able to find a solution? Click here to request help.