Gradient Start Delay does not work on H-Class system with ISM and QDa - WKB8884
Article number: 8884
SYMPTOMS
- Gradient Start Delay set in the instrument method (for QSM) does not function as expected; whether setting it before or after injection, or specifying a volume vs. specifying a time
ENVIRONMENT
- ACQUITY UPLC H-Class system comprising:
- QSM
- SM-FTN
- CM
- PDA
- QDa
- ISM
CAUSE
ACQ-ISM set as the Primary Flow Source in the ACQUITY Console. This setting is found by highlighting the system in the left-hand pane of the ACQUITY Console and then selecting from the Configure menu > Primary Flow Source.
FIX or WORKAROUND
In the ACQUITY Console, change the Primary Flow Source setting to the ACQ-QSM.
ADDITIONAL INFORMATION
id8884, QDA, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO