ACQUITY H-Class high RSD and variable system pressure - WKB88843

Article number: 88843

SYMPTOMS

  • High %RSD
  • Variable pressure
  • High delta pressure

ENVIRONMENT

ACQUITY H-Class

CAUSE

Improperly swaged APH fitting at port 6 of the SM-FTN injector pod.

FIX or WORKAROUND

Replace the APH assembly (APH, SS, 12.5-in - 205001774 or APH, MP35N, 12.5-in - 205001754)