Broad peak shape with Alliance HPLC System - WKB88620

Article number: 88620

SYMPTOMS

  • Peak shape approximately twice as broad, 60% sensitivity
  • Static leak test, valve leak tests, adjust seals and compression check pass

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

Pre-column tubing is damaged

FIX or WORKAROUND

Replace the pre-column tubing.

or

Using a tubing cutter, cut the tubing back and replace the compression screw and ferrule.

ADDITIONAL INFORMATION

WAT025313 - Compression screw

WAT022330 - Ferrule

WAT026973 - 0.009 x 1/16 x 10 ft SS tubing