Column manager outlet tubing is blocked - WKB88534
Article number: 88534
SYMPTOMS
- The tubing to all column manager positions pops out when flow is applied to the system
ENVIRONMENT
- ACQUITY UPLC Column Manager
CAUSE
The outlet tubing from the column manager pod to the MS/PDA or detector is blocked, causing back pressure to all of the CM column positions.
FIX or WORKAROUND
- Replace the column manager outlet tubing.
- Apply flow to the system and inspect for leaks in any of the tubing.
- If none are present, ensure flow is observed from the end of the outlet tubing to the MS/PDA .
ADDITIONAL INFORMATION