ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM) - Retention time shifts with tendency (earlier elution) - WKB88528
Article number: 88528
SYMPTOMS
- Retention time shifts with trend toward earlier peak elution.
- Subsequent injection shows earlier retention time (difference of 0.1-0.2 sec).
ENVIRONMENT
ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
CAUSE
Blocked GPV low-pressure inline filters (700005173).
FIX or WORKAROUND
- Replace GPV low-pressure filters (700005173).
- Disconnect GPV to mixer tube fittings.

- Take off old GPV low-pressure inline filters.

- Replace with new GPV low-pressure inline filters.

- Disconnect GPV to mixer tube fittings.
- Prime system.
- Run test injections to confirm RT shift resolution.