Skip to main content
Waters

Syringe fails to move - WKB88523

Article number: 88523

SYMPTOMS

  • Sample syringe fails to move on startup
  • Plunger error
  • 2767: Syringe: > Invalid command

ENVIRONMENT

  • 2767 Sample Manager

CAUSE

Failed syringe drive motor or board

FIX or WORKAROUND

Replace the syringe drive motor and board (700003531).

ADDITIONAL INFORMATION

 

id88523, 2767, 2767SFC