Data buffer overflow error on ACQUITY UPLC TUV in Console - WKB88452

Article number: 88452

SYMPTOMS

  • Data buffer overflow error on TUV displayed in the instrument console.
  • TUV was unable to change wavelength or view its system info without crashing the console.

ENVIRONMENT

  • ACQUITY UPLC TUV
  • Standalone Console

CAUSE

Bad communications between the TUV and the PC. The DHCP server only displayed a couple of the system modules and not all of their info.

FIX or WORKAROUND

  1. Power-off modules, remove all the present modules from the DHCP, run Stop ACQUITY Processes, restart the PC. Power-on the Ethernet hub containing the module first, usually the sample manager. Ensure that it correctly populates the DHCP server, and then power-on the remaining modules. If the DHCP contains all modules, then open the Console and confirm the performance
  2. If poor communication persists or a module does not populate the DHCP, consider replacing its CPU. 

ADDITIONAL INFORMATION

 

comms, UPBINARY, UPPTUV, UPSMPMGR, UPTUV, UPTUVARC, UPTUVARCB, UPTUV-E, UPTUVTC