Article number: 88366

OBJECTIVE or GOAL

Set the valve shaft in the correct position in order to replace the Everflow injector valve POD.

ENVIRONMENT

  • 2777/2777c Sample Manager

PROCEDURE

1. Using the 2777/2777C keypad, press F1 to select the Menu function key.

2. Spin the outer knob clockwise to scroll to Utilities, and then press Enter.

3. Scroll to Injector, and then press Enter.

4. Scroll to select LCVlv1, and then press Enter.

5. Press F2 (ACT Valve) to set the correct valve drive shaft position.

ADDITIONAL INFORMATION