With 2695 module and 2489 UV/Vis detector, peak area of the target component varies, but the peak area of the simultaneously measured IS does not - WKB87731
Article number: 87731
SYMPTOMS
- The peak area of the target component varies, even though the peak area of the simultaneously measured IS does not
- No problem with the retention time and the peak shape
- Mobile phase contains ion-pairing reagent and is premixed
- No sample preparation issues
ENVIRONMENT
- 2695 Separations Module
- 2489 UV/Vis Detector
CAUSE
Column condition changes due to continued use with the same mobile phase composition.
FIX or WORKAROUND
- If you have another column of the same type, use it to measure and see if it solves the problem.
- If there is no other column of the same type, try cleaning the column.
ADDITIONAL INFORMATION
Carryover solution that can be performed by user with 2695 Separations Module
