With 2695 module and 2489 UV/Vis detector, peak area of the target component varies, but the peak area of the simultaneously measured IS does not - WKB87731
Article number: 87731
SYMPTOMS
- The peak area of the target component varies, even though the peak area of the simultaneously measured IS does not
- No problem with the retention time and the peak shape
- Mobile phase contains ion-pairing reagent and is premixed
- No sample preparation issues
ENVIRONMENT
- 2695 Separations Module
- 2489 UV/Vis Detector
CAUSE
Column condition changes due to continued use with the same mobile phase composition.
FIX or WORKAROUND
- If you have another column of the same type, use it to measure and see if it solves the problem.
- If there is no other column of the same type, try cleaning the column.
ADDITIONAL INFORMATION
Carryover solution that can be performed by user with 2695 Separations Module
id87731, 2489, 2489B, 2489SFC, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, A-10UV, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, eluent