ACQUITY I-Class SM-FL error: "Sample pressure low" - WKB86781

Article number: 86781

SYMPTOMS

  • "Sample pressure low (##): possibly no wash solvent" error appears when trying to perform an injection.
  • Sample and wash syringe leak tests pass.
  • Needle seal leak test passes.
  • Back pressure regulator passes.
  • Needle seal and needle loop and volume characterizations pass.
  • No visible leaks at wash manifold.

ENVIRONMENT

  • ACQUITY UPLC Sample Manager
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC I-Class PLUS
  • ACQUITY UPLC M-Class
  • nanoACQUITY UPLC

CAUSE

PEEK connector that connects the line that carries wash solution from the wash manifold to the SM/SM-FL needle wash station is lose or damaged.

FIX or WORKAROUND

Reconnect, tighten, or replace the fitting that connects the wash line to the needle wash station. 

ADDITIONAL INFORMATION