No pressure on ACQUITY UPLC SM-FTN when injection valve is in inject position - WKB86773

Article number: 86773

SYMPTOMS

  • No pressure when FTN sample manager is in inject position
  • Normal pressure when in load position
  • Solvent Manager (QSM) leak tests pass

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class
  • ACQUITY Arc SM-FTN

CAUSE

The sample needle is not properly positioned in the needle carriage and does not seal on the needle seal when the SM-FTN is in inject position.

FIX or WORKAROUND

  1. Inspect for damage to the needle and seal and replace as necessary.
  2. Replace the needle and position it correctly in the carriage. Consult the "how to" video for reference: https://videos.waters.com/detail/videos/acquity-how-to-series/video/82090424001/replacing-the-needle-guide-and-vespel-seat?autoStart=true

ADDITIONAL INFORMATION

 