How to replace the ACQUITY UPLC Column Manager switching-valve cartridge. - WKB85242
Article number: 85242
OBJECTIVE or GOAL
Replacing the column manager switching-valve cartridge.
ENVIRONMENT
- ACQUITY UPLC Column Manager
-
ACQUITY UPLC H-Class PLUS
-
ACQUITY UPLC H-Class PLUS Bio
-
ACQUITY UPLC I-Class PLUS
PROCEDURE
- In the console, select Column Manager from the system tree and click Interactive Display.
- In the column manager information window, click Column Selection.
- In the Column Manager Set Column dialog box, select Waste *
- Pull the valve forward, approximately 2 cm, and rotate the lock ring clockwise until it rests against the lock ring stop.
- Use the 1/4-inch wrench to remove the fittings attached to the switching-valve cartridge.
- Rotate the lock ring counterclockwise and use the 2-mm hex wrench to remove the hex
screw at the 10 o’clock position on the switching-valve cartridge. - Remove the switching-valve cartridge from the switching-valve assembly by pulling the
cartridge straight forward. - Unpack the new switching-valve cartridge.
- Ensure that the groove in the cartridge housing aligns with the groove on the drive clamp.
- Insert the new switching-valve cartridge into the switching-valve cartridge chamber.
- Insert the 2-mm hex screw at the 10 o’clock position on the switching-valve cartridge and
tighten the screw using the 2-mm hex wrench. - Pull the valve forward, approximately 2 cm, and rotate the lock ring clockwise until it rests
against the stopper. - Use the 1/4-inch wrench to reattach all fittings and tighten them 1/4-turn beyond finger-tight
for existing fittings, or 3/4-turn beyond finger-tight for new fittings. - Rotate the lock ring counterclockwise to the 12 o’clock position.
- Push the switching-valve assembly in until it stops.
- Ensure that the tubing is properly oriented so that the column compartment door is
unobstructed when closed.
*
NOTE: If the column manager has been configured with Advanced settings eg use as a diverter valve, the Waste will not be an option therefore select Position 2 (waste)
ADDITIONAL INFORMATION
Refer to page 93 of the ACQUITY UPLC Column Manager - Active (CM-A) and Column Manager - Auxiliary (CM-Aux) for further information.