ACQUITY UPLC H-Class shows "APH active but not present" error - WKB83947
Article number: 83947
SYMPTOMS
- "APH active but not present" error seen for one column position
- LC runs do not start
ENVIRONMENT
- UNIFI Network
- ACQUITY UPLC H-Class
- ACQUITY UPLC Column Manager
CAUSE
A 30-cm column oven and a column manager are installed. Column manager is configured in 4-column mode but has only two APH installed. Position 4 from the column manager inlet pod runs to the 30-cm column manager. APH for positon 4 is set to active.
FIX or WORKAROUND
- Set APH 4 to "Present Passive".
- Set APH 4 to "Present Passive" in the Column manager section of the LC method.
ADDITIONAL INFORMATION
APH and column temperature for the 30-cm column oven are set in the sample manager console page. The column manager has only two column positions plumbed. The outlet for position 4 is plumbed to the 30-cm column oven. The column manager must be configured for four positions to allow for column 4 to be selected. This allows flow to the 30-cm column oven from the column manager.