SYMPTOMS
- Wash solvent bottle empties rapidly even when system is off.
- A leak appears on the drip tray under the syringe.
ENVIRONMENT
-
ACQUITY UPLC H-Class SM-FTN
-
ACQUITY UPLC H-Class Bio SM-FTN
-
ACQUITY UPLC H-Class PLUS SM-FTN
-
ACQUITY UPLC H-Class PLUS Bio SM-FTN
-
ACQUITY UPLC I-Class SM-FTN
-
ACQUITY UPLC I-Class PLUS SM-FTN
CAUSE
The ferrule that fixes the wash solvent tube to the pump is loose.
FIX or WORKAROUND
- Remove the FTN right panel.
- Tighten the ferrule that fixes the wash solvent tube to the pump.
ADDITIONAL INFORMATION
id83852, UPARHSMFTN