How to shut down the ACQUITY UPLC ELS Detector - WKB83139
Article number: 83139
OBJECTIVE or GOAL
Shut down the ELS detector for a holiday period.
ENVIRONMENT
- MassLynx
- ACQUITY UPLC ELS Detector
PROCEDURE
- Remove all buffers from the column and detector by flowing an unbuffered mobile phase through the system.
- Set the pump to OFF.
- Allow the nebulization gas to flow through the detector for ten minutes to drain the evaporation tube and detection chamber.
- Stop the gas flow.
- Power-off the detector.
ADDITIONAL INFORMATION
For more information, see the ACQUITY UPLC Evaporative Light Scattering Detector Getting Started Guide (71500109303).