3777c syringe leaks - WKB83102

Article number: 83102

SYMPTOMS

  • Liquid collects on the top of the syringe barrel
  • Liquid pools at the base of the syringe

ENVIRONMENT

  • 3777c Sample Manager

CAUSE

The syringe is loose or worn.  

FIX or WORKAROUND

Tighten the syringe, or replace the syringe (700011725).

ADDITIONAL INFORMATION

leaking syringe_2.jpgleaking syringe_1.jpg