ACQUITY RI noisy baseline - WKB83098
Article number: 83098
SYMPTOMS
- During SQT, baseline is wavy. Noise and drift test pass, but peak integration is not consistent.
- The diluent peak is showing as a positive peak instead of the usual negative peak.
- Occurring in both standard and samples.
ENVIRONMENT
- 2414 Refractive Index Detector
- ACQUITY UPLC RI Detector
CAUSE
Dirty flow cell.
FIX or WORKAROUND
- Purge detector with 30% phosphoric acid, followed by water.
ADDITIONAL INFORMATION
id83098, 2414, 2414B, A-10RI, A-30RI, UPRI, UPRIARC