Skip to main content
Waters

ACQUITY RI noisy baseline - WKB83098

Article number: 83098

SYMPTOMS

  • During SQT, baseline is wavy. Noise and drift test pass, but peak integration is not consistent.
  • The diluent peak is showing as a positive peak instead of the usual negative peak.
  • Occurring in both standard and samples.

ENVIRONMENT

  • 2414 Refractive Index Detector
  • ACQUITY UPLC RI Detector

CAUSE

Dirty flow cell.    

FIX or WORKAROUND

  1. Purge detector with 30% phosphoric acid, followed by water.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.