What is the dwell volume for an ACQUITY UPLC H-Class system? - WKB8302
Article number: 8302
ENVIRONMENT
- ACQUITY UPLC H-Class System
- ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
- ACQUITY UPLC
- H-Class
- PLUS
- QSM
- QSM+
ANSWER
Dwell volume, system is <400 μL with 100-μL mixer.
ADDITIONAL INFORMATION
With the Radial GPV Kit, part number 205001661 installed in a H-Class System the Dwell Volume Specification will increase to < 505 uL.
Dwell Volume and Extra-Column Volume: What Are They and How Do They Impact Method Transfer ACQUITY UPLC H-Class System Guide.
id8302, A-30SDM, gradient proportioning valve, UPAQSM, UPBSMIVD, UPIBSMIVD, UPLCCOLHTR, UPNBINARY, UPPQSM, UPQSM, UPQSM, UPQSM+, UPQSM+, UPQSMBIO, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMIVD, UPSMPMGR