UPLC pump overpressure - WKB82448

Article number: 82448

SYMPTOMS

  • UPLC overpressure error on pump        

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC I-Class

CAUSE

Partial blockage at column tubing on position 6 of injection pod    

FIX or WORKAROUND

  1. Remove and refit the column tubing at port 6 on the injection pod.
  2. Replace the tubing if required.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.