BSM is unable to compress during leak test (second issue) - WKB82367

Article number: 82367

SYMPTOMS

  • BSM is unable to compress on accumulator pump during leak test

ENVIRONMENT

  • ACQUITY BSM

CAUSE

A fitting is leaking on the steel tubing between the pump head and the pressure transducer.

FIX or WORKAROUND

Replace the steel tubing (430001120).

ADDITIONAL INFORMATION

The transducer to check valve tubing is also a possible cause and may need replacing (I-Class: 430002583, Classic Acquity i2V: 430001773, Classic Acquity Passive: 430001121).