Alliance 2695 error: Needle malfunction - WKB82326
Article number: 82326
SYMPTOMS
- Error message during initialization: "Needle malfunction"
ENVIRONMENT
- 2695 Separations Module
- Alliance HPLC system
CAUSE
Dust on the block stems for an old style injector

FIX or WORKAROUND
Clean the injector block stems with methanol, and use WD40 if necessary.
ADDITIONAL INFORMATION
